SmartReader Data Loggers
34
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
Thermocouple Type
Positive (+) Wire
Negative (-) Wire
J
White
Red
K
Yellow
Red
T
Blue
Red
S
Black
Red
2. Connect the logger to your computer and run TrendReader. Select the equation number
for each active thermocouple channel (refer to the Specifications section).
3. Check that you have the thermocouple's polarity correct by warming its tip. You will see
the Realtime temperature reading increase if the wires are connected correctly. If the
temperature decreases, reverse the connections.
NOTE: The inputs are not isolated from each other (they use the same common
"-" terminal internally). This means that the thermocouple tips cannot be touching
each other. If they are touching or are fastened to a conducting surface like a
metal frame, some looping can occur and the readings may fluctuate. They must
be insulated or prevented from making contact with each other electrically. Do
not use grounded thermocouples.
1.8.3 Specifications
Accuracy: ±1% of range + resolution
Type J Thermocouple Inputs
Thermocouple
Equation
Range
Resolution
1 (Narrow)
49
-20 to 190°C
(0 to 370°F)
1.4°C (2.5°F)
2 (Wide)
50
-50 to 600°C
(-55 to 1100°F)
5°C (9°F)
Type K Thermocouple Inputs
Thermocouple
Equation
Range
Resolution
1 (Narrow)
51
-25 to 230°C
(-10 to 440°F)
1.7°C (3°F)
2 (Wide)
52
-100 to 900°C
(-145 to 1650°F)
6.7°C (12°F)
Type S Thermocouple Inputs
Thermocouple
Equation
Range
Resolution
1 (Narrow)
73
0 to 960°C
7°C (12.6°F)