II-3.
FLUOROSCOPIC TIMER
A. Five Minute Maximum Setting
B. Audible Warning Functional
C. Accuracy of Elapsed Exposure time (at least
annually or after repairs or replacement)
II-4.
SHIELDING DEVICES AND APPAREL
A.
B.
C.
D.
E.
F.
H.
Lead Drape
1. Presence
2. Condition
Bucky Slot Shield
1. Presence
2. Working condition
3. Adequacy
Table End shield Presence
Aprons (See Section I-4-A)
1. Available
2. Employed
Gloves (See Section I-4-B)
1. Available
2. Employed
Gonadal
Shield (See Section I-4-C)
1. Available
2. Employed
G. Mobile Shield (See Section I-4-D)
Other Specialized Devices
II-5.
MINIMUM SOURCE-TO-SKIN DISTANCE (SSD)
A. Minimum Fluoroscopic SSD
1. X-ray tube below table (after a change
in type or mounting)
2. X-ray tube above table
3. C-arm
II-6.
RADIATION BEAM RESTRICTION AND ALIGNMENT
(Sizing checks should be performed at least an-
nually or
after repairs
or adjustment to the
collimator, its electronics, image intensifier
tower SID sensing device, or cassette holder
within
spotfilm
device.
Alignment checks
should be performed at least annually or after
adjustments or repairs to the cassette holder
of the spot film device, to the photospot camera,
to the television camera, or to the television
monitor.)
A. Fluoroscopic Mode Alignment and Sizing
(Minimum SID, auto and/or manual mode)
1. Large FOV (14", 13", 12", 11", 10", or 9")
2. Medium FOV (10", 9", 7", or 6")
3. Small FOV (6", 5", or 4 l/2")
35